SHS-40
Automated Headspace Sampler
Providing Flexbility, Performance, and Value
Innovation with Integrity
Gas Chromatography
Getting the Most From Your GC System The SHS-40 Automated Headspace Sampler is designed to take maximum advantage of the productivity and cost benefits of Headspace – Gas Chromatography (HS-GC). Powerful and robust, the SHS-40 enhances the capabilities of any analytical laboratory to comprehensively and efficiently measure volatile organic compounds (VOCs) from a variety of sources. Specifically engineered with performance and value in mind, the SHS-40 provides a number of benefits that make it an ideal component of a GC-MS system for use in routine VOC analyses.
Fully Automated Minimal Operator Intervention Highly Productive Simplified Method Development Easily Integrated Ready to Use with Bruker GC and GC-MS Systems Flexible Use for Direct Analysis or “React and Analyze” Chemistries Robust Very Low Maintenance Requirements
The Flexibility and Performance to Do More The SHS-40 is the analyst’s choice for headspace sampling and analysis. Designed to match your most stringent analytical needs, it incorporates many exceptional features to maximize performance and flexibility. The SHS-40 is available in two forms: System Features
Fixed Sample Volume Configuration
Sample Capacity
Up to 40 vials; 20mL or 10 mL w/adaptor Crimp or Screw Top Vials
Injection Modes
Fixed Sample Volume Press and Inject
Injection Valve
6 Way Electric Actuated 1 mL Sample Loop Heat up to 350°C
Incubator
Up to 12 Samples 40-200°C in 1°C increments Integrated Shaker
Instrument Control
Remote with Compass CDS Stand alone
Sample Recognition and Detection
Automated
Carrier Gas Control
Direct from GC
Optional Accessories
Additional Sampling Valve Reactant Gas Catalytic Converter
SHS-40
Press and Inject Configuration
SHS-40
The
Bringing Performance and Value to a Range of Applications The SHS-40 is widely deployed in a number of key analytical sectors ranging from Pharmaceutical, Flavor/Fragrance, Petrochemical/Energy, Forensic/Toxicology, Polymers/ Packaging and Environmental Testing laboratories.
TOGA Gas Analysis CO2
N2
O2
CO
Residual Solvents Analysis Residual solvents in pharmaceuticals are defined as organic volatile chemicals that are consumed or produced in the manufacture of drug products that are not completely removed by practical manufacturing techniques. The United States Pharmacopeia (USP) specifies the gas chromatographic (GC) conditions for the analysis of these organic volatile impurities. The SHS-40 and Bruker GC combine to provide excellent results for this method. Residual Solvent Peak Analysis
C2H4
CH4
C2H2
C2H6
H2
0
1
C3H6
2
3 Min
0
5
TCD Channel
10
15
Min
FID Channel
Biodiesel – Methanol Analysis In order for biodiesel to be used as a motor fuel or blended with petroleum diesel, it must conform to standard EN-14110 to determine methanol content. Utilization of the Bruker SHS-40 and GC combination, allows for the application of this method to biodiesel samples. The system generates excellent results that confirm that the HS-GC system is well suited for the determination of methanol content in biodiesel according to specifications outlined in EN-14110. Typical Headspace Chromatogram of Biodiesel
Transformer Oil Gas Analysis Insulating fluids, generally mineral oils, are often used in electrical transformers. Occasionally, localized or general heating of the oil occurs and decomposition products are formed. If the concentration of these gases reaches a critical point, the chances of catastrophic transformer failure are high. The ASTM D 3612 method 1 (Part B) describes in detail an HS-GC method for the measurement of these gases. The analysis of dissolved gases in transformer oil according to ASTM D3612, method B, can be performed easily and effectively with the Bruker TOGA and SHS-40 headspace sampler. Using a Bruker system results in full separation of all components of interest with easy and reliable quantitation.
20
Methanol Calibration Curve
25
Simple and Productive Integration with GC-MS
Headspace with Gas Chromatography – Mass Spectrometry (HS-GC-MS) is an extremely powerful technique to characterize and measure VOC’s. Utilizing this technique is ideal for the analysis and quantitation of known and well characterised components. Taking advantage of the potent combination of analytical platforms couldn’t be easier with the Bruker SHS-40. Ready made for integration with the Bruker 400 Series GC and powerful SCION Series GC-MS platforms, the SHS-40 fits in perfectly with these companion systems to create the ultimate platform for VOC analysis.
Mass Spectrum and Library Search of VOCs with A Fully Integrated Bruker HS-GC-MS System
System Specifications • Sample Capacity: 40 x 20ml (10ml vials with adaptors). Fully integrated, automated vial queuing system • Automated Gas Sampling Valve- Electric actuated 6-way injection valve (VICI) with 1 ml sample loop. This entire module may be separately heated up to 350°C. • Heated incubation oven for 12 samples (40°C - 200°C, in 1°C-steps • Integrated Shaker • Automatic sample detection/recognition • Compatible with Crimp and Screw Top Vials • Carrier gas controlled direct from the GC • AUX gas pre-set internally. • Up to 9 different parameter sets can be stored. • Adaptable to different injectors • Multiple Headspace Extraction mode via single puncture • Flexible transfer line • Option: Sample Transfer Line Kit (direct connection to carrier gas, insertion of transfer line via injector no longer required) • Control: Stand alone or remotely via Compass CDS software • Voltage: 110V or 230V • Dimensions: Width: 29cm (12in) Height: 46cm (18in) Depth: 62cm (25in) • Weight: 21kg (47lb)
Billerica, MA · USA Phone +1 (978) 663-3660 Fax +1 (978) 667-5993
[email protected]
www.bruker.com
Fremont, CA · USA Phone +1 (510) 683-4300 Fax +1 (510) 490-6586
[email protected]
Bremen · Germany Phone +49 (0)421-2205-0 Fax +49 (0)421-2205-103
[email protected]
the right to change specifications without notice. © BDAL 06-2011
Bruker Daltonik GmbH
Bruker Daltonics Inc.
Bruker Daltonics is continually improving its products and reserves
For research use only. Not for use in diagnostic procedures.